IRt/c.01
Sensing Range -50 to 550 °F (-45 to 290 °C) Ambient Temperature Range 0 to 160 °F (-18 to 70 °C) Optimum Range Selections Eight models per t/c type Field-of-View 1:1 (60°) approximately Minimum Spot Size 0.3" (8 mm) Spectral Response 6.5 to 14 µ Output Impedance 3 Kohms approximately Cable Thermocouple extension grade, PVC jacket, unshielded Dimensions 1.28" x .71" Diameter (32.5 x 20 mm) Weight 1.4 oz (40 g) with cable Housing High strength ABS, hermetically sealed, exceeds NEMA4,4x; IP65, 67, intrinsically safe. Two locknuts supplied. |
|
IRt/c.03
Sensing Range -50 to 550 °F (-45 to 290 °C) Ambient Temperature Range 0 to 160 °F (-18 to 70 °C) Optimum Range Selections Eight models per t/c type Field-of-View 3:1 (17°) approximately Minimum Spot Size 0.25" (6 mm) Spectral Response 6.5 to 14 µ Output Impedance 4 to 8 Kohms approximately Cable Thermocouple extension grade, PVC jacket, unshielded Dimensions 1.28" x .71" Diameter (32.5 x 20 mm) Weight 1.4 oz (40 g) with cable Housing High strength ABS, hermetically sealed, exceeds NEMA 4, 4x; IP65, 67, intrinsically safe. Two locknuts supplied. |


|
IRt/c
Sensing Range -50 to 1200 °F (-45 to 650 °C) Ambient Temperature Range 0 to 212 °F (-18 to 100 °C) Optimum Range Selections Eight models per t/c type Field-of-View 1:1 (60°) approximately Minimum Spot Size 0.3" (8 mm) Spectral Response 6.5 to 14 µ Output Impedance 3 Kohms approximately Cable Twisted shielded pair of base thermocouple material (J, K, etc.), 3 ft. (.9) m) standard length, Teflon sheathed, rated to 392 ºF (200 ºC) continuous service. Dimensions 1.75" x .50" Diameter (44 x 12.7 mm) Weight 1.4 oz (40 g) with cable Housing Stainless steel, hermetically sealed, exceeds NEMA 4, 4x; IP65, 67, intrinsically safe, cable shield grounded to housing and electrically isolated from signal. |
|
IRt/c.SV
Sensing Range -50 to 1200 °F (-45 to 650 °C) Ambient Temperature Range 0 to 212 °F (-18 to 100 °C) Optimum Range Selections Eight models per t/c type Field-of-View 1:1 (60°) approximately Minimum Spot Size 0.3" (8 mm) Spectral Response 6.5 to 14 µ Output Impedance 3 Kohms approximately Cable Twisted shielded pair of base thermocouple material (J, K, etc.), 3 ft. (.9) m) standard length, Teflon sheathed, rated to 392 ºF (200 ºC) continuous service. Dimensions 2.27" x .50" Diameter (57.8 x 12.7 mm) Weight 1.4 oz (40 g) with cable Housing Stainless steel, hermetically sealed, exceeds NEMA 4, 4x; IP65, 67, intrinsically safe, cable shield grounded to housing and electrically isolated from signal. |

|
IRt/c.1X
Sensing Range -50 to 1200 °F (-45 to 650 °C) Ambient Temperature Range 0 to 212 °F (-18 to 100 °C) Optimum Range Selections Eight models per t/c type Field-of-View 1:1 (60°) approximately Minimum Spot Size 0.3" (8 mm) Spectral Response 6.5 to 14 µ Output Impedance 3 Kohms approximately Cable Twisted shielded pair of base thermocouple material (J, K, etc.), 3 ft. (.9) m) standard length, Teflon sheathed, rated to 392 ºF (200 ºC) continuous service. Dimensions 1.43" x .50" Diameter (36 x 12.7 mm) Weight 1.4 oz (40 g) with cable Housing Stainless steel, hermetically sealed, exceeds NEMA 4, 4x; IP65, 67, intrinsically safe, cable shield grounded to housing and electrically isolated from signal. |


|
IRt/c.3X
Sensing Range -50 to 1200 °F (-45 to 650 °C) Optimum Range Selections Eight models per t/c type Field-of-View 3:1 (17°) approximately Minimum Spot Size 0.25" (6 mm) Spectral Response 6.5 to 14 µ (2 to 20 µ with IRt/c.3XG germanium lens) Output Impedance 4 to 8 Kohms approximately Cable Twisted shielded pair of base thermocouple material (J, K, etc.), 3 ft. (.9) m) standard length, Teflon sheathed, rated to 392 ºF (200 ºC) continuous service. Dimensions 1.76" x .50" Diameter (44.7 x 12.7 mm) Weight 1.4 oz (40 g) with cable Housing Stainless steel, hermetically sealed, exceeds NEMA 4, 4x; IP65, 67, intrinsically safe, cable shield grounded to housing and electrically isolated from signal. Air Purge Built-in; designed for severe paint or ink environment; 3; (0.9 m) of tubing provided. |
|
IRt/c.3SV
Sensing Range -50 to 1200 °F (-45 to 650 °C) Optimum Range Selections Eight models per t/c type Field-of-View 3:1 (17°) approximately Minimum Spot Size 0.2" (5 mm) Spectral Response 6.5 to 14 µ Output Impedance 4 to 8 Kohms approximately Cable Twisted shielded pair of base thermocouple material (J, K, etc.), 3 ft. (.9) m) standard length, Teflon sheathed, rated to 392 ºF (200 ºC) continuous service. Dimensions 2.27" x .50" Diameter (44.7 x 12.7 mm) Weight 1.6 oz (44 g) with cable Housing Stainless steel, hermetically sealed, exceeds NEMA 4, 4x; IP65, 67, intrinsically safe, cable shield grounded to housing and electrically isolated from signal. Air Purge Built-in; designed for severe paint or ink environment; 3; (0.9 m) of tubing provided |

|
IRt/c.5
Sensing Range -50 to 1200 °F (-45 to 650 °C) Optimum Range Selections Eight models per t/c type Field-of-View 5:1 (11°) approximately Minimum Spot Size 0.8" (20 mm) Spectral Response 6.5 to 14 µ Output Impedance 4 to 8 Kohms approximately Cable Twisted shielded pair of base thermocouple material (J, K, etc.), 3 ft. (.9) m) standard length, Teflon sheathed, rated to 392 ºF (200 ºC) continuous service. Dimensions 3.4" x 1.375" Diameter (86 x 35 mm) Weight 6.5 oz (184 g) with cable Housing Stainless steel, hermetically sealed, exceeds NEMA 4, 4x; IP65, 67, intrinsically safe, cable shield grounded to housing and electrically isolated from signal. Air Purge Built-in; cooling capacity to 400 °F (200 °C) ambient; 3; (0.9 m) polyurethane tubing provided |

|
IRt/c.10
Sensing Range -50 to 1200 °F (-45 to 650 °C) Optimum Range Selections Eight models per t/c type Field-of-View 10:1 (6°) approximately Minimum Spot Size 0.8" (20 mm) Spectral Response 6.5 to 14 µ Output Impedance 4 to 8 Kohms approximately Cable Twisted shielded pair of base thermocouple material (J, K, etc.), 3 ft. (.9) m) standard length, Teflon sheathed, rated to 392 ºF (200 ºC) continuous service. Dimensions 3.76" x 1.375" Diameter (96 x 35 mm) Weight 6.5 oz (184 g) with cable Housing Stainless steel, hermetically sealed, exceeds NEMA 4, 4x; IP65, 67, intrinsically safe, cable shield grounded to housing and electrically isolated from signal. Air Purge Built-in; cooling capacity to 400 °F (200 °C) ambient; 3; (0.9 m) polyurethane tubing provided |

|
Selecting a Pre-Calibrated IRt/c
Ordering information: (Model) - (Thermocouple Type) - (Pre-Calibrated Temperature Range) Select IRt/c(*) model. Example: IRt/c.3X Select the thermocouple type (**) desired (J, K, etc.), add it to the model name. Example: IRt/c.3X-K Select the target temperature range. For example, to control a lamination process at 200 F (93 C), look at the following target temperature table for 200 F (93 C). Note the black area in the table indicates the "180F/90C" pre-calibrated range. Add this to complete your selection. Example: IRt/c.3X - K - 180F/90C. Additional note: The above table indicates the temperature ranges where the IRt/c’s match standard thermo-couple signals. This allows for highly accurate measurement and control in the ranges selected. IRt/c’s require no power supply and can be connected directly to thermocouple inputs of controllers, PLC’s, transmitters, and other t/c devices. In addition, all IRt/c sensors are useable over the entire specified target temperature range for each sensor. Signal output tables and polynomials are available from Exergen. Accuracy IRt/c’s are pre-calibrated at the factory for typical target material emissive properties, but actual emissivi-ties may differ. The accuracies specified assume that the precalibrated value is correct for your installation, or that the single point set up of Tech Note #1 has been performed to reduce the error to zero at the calibra-tion point (which is normally the set point for a temperature control system). |

|
If you can't find what you are looking for
Call Us! or go to the
Exergen Home Page |
|
Seltrol is a stocking distributor of a wide
variety of industrial automation equipment. We are located in South
Carolina and service SC and NC. Our inside and outside sales staff are
available to assist you with your application. We provide technical -
application support for all of our products.
|
|
Exergen Remote IR Temperature Sensors
|